Description: Used in good condition as shown in the pictures. Tested and working. We used an LA435 linear transducer in our testing as shown. No transducers are included with the ultrasound. This sale is for the MyLab 40 ultrasound system only as shown. The ultrasound is in clean cosmetic condition with no damage except three of the little plastic hooks next to the transducer holders are broken as shown in the pictures. The only issue we noted is when shutting the ultrasound down, after you press the power button located under the screen it goes to a screen that states "Please wait; the system is shutting down" as shiwn in the kast picture. We left it running for 30 minutes waiting and it never left this screen or turned off so we turned the main power off located on the back top left of the ultrasound. Then just to check we restarted the ultrasound and everything started up fine, when we pressed the power button to shut it down again, it did the same thing. We tested it again the same way. Every time, it stated on the screen stating "Please wait; the system is shutting down" and after waiting several minutes, we had to turn the power off using the main power switch. I believe these are supposed to turn off after it finishes the shut down process, but this one doesn't so I would recommend when you shut it down, wait several minutes to allow it to complete its shutdown procedure before turning off the main power switch. I do not have a manual for this ultrasound. Item#923130L